Separation of two model proteins with Ion Exchange Chromatography (IEC) Michaela Schulze, Ulrike Maschke, Kate Monks; [email protected] KNAUER Wissenschaftliche Geräte GmbH, Hegauer Weg 38, 14163 Berlin; www.knauer.net SUMMARY VBS0065 © KNAUER Wissenschaftliche Geräte GmbH Additional information Ion exchange chromatography is one of the most widely used FPLC techniques for protein separation and purification. This application describes an easy separation of two model proteins and explains how ion exchange chromatography works. INTRODUCTION Ion exchange chromatography (IEC) separates molecules according to type and strength of their charge. The isoelectric point (pI) is the pH where a protein or molecule has no net elec- trical charge. Depending on the pH of the buffer a protein has different surface charges in solution. At a pH above their pI proteins have a negative charge and bind to positively charged resins such as anion exchangers. At a pH below their pI proteins have positive charge and bind to negatively charged cation exchangers. This interaction is used for the separation and purification of various proteins. By using a suitable pH and low salt conditions proteins bind to the resin in the initial step ( Fig. 1A). The proteins are mostly separated with a linear salt gradient. The salt ions compete with the proteins for bindings sites. Proteins with weak ionic interactions are the first to elute from the column. In the case of cation exchange chromato- graphy, proteins that are less negatively charged start to elute first ( Fig. 1B ). With an increase of the salt concentration proteins with stronger ionic interaction elute later from the column ( Fig. 1C). RESULTS α-Chymotrypsinogen A (pI 8.97) and Lysozyme (pI 11.35) are proteins with rather high pI values, which make them ideal candidates for cation exchange chromatography ( Fig. 2 ). Both proteins bind under low salt conditions to the resin. During the wash step that removes potentially unbound protein only a small amount of impurities eluted from the column. α-Chymot- rypsinogen A eluted first from the column due to its lower pI of 8.97 ( Fig. 2, peak 1). With an increasing gradient and therefore increasing salt concentration lysozyme eluted as second peak ( Fig. 2, peak 2). The eluted proteins were automatically collected by the fraction collector. The salt gradient was monitored by the conductivity monitor ( Fig. 2, red signal). MATERIALS AND METHOD In this application, an AZURA® Bio purification system consisting of AZURA P 6.1L LPG metal-free pump with 10 mL pump head; AZURA ASM 2.1L assistant module with feed pump and two injection valves; an AZURA DAD 2.1L diode array detector with 10 mm, 10 µL flow cell; AZURA CM 2.1S conductivity monitor and Foxy R1 fraction collector was used. Prior to the run the 1 mL cation exchange column (HiTrap Capto S) was equilibrated with 10 mL buffer A (20 mM phosphate buffer pH 7.1). The flowrate for the run was 2 mL/min. 500 µL protein mixture (α-Chymotrypsinogen A 2.5 mg/ mL + Lysozyme 2.5 mg/mL) was injected. Subsequently the column was washed with 4 mL buffer A to remove all unbound protein. The two proteins were eluted with a linear gradient over a total volume of 70 mL. The salt concentration increased with the rising concentraion of buffer B (20 mM phosphate buffer pH 7.1 + 1M NaCl) in the gradient from 0 % up to 30 %. Eluting proteins were collected with the fraction collector. The column was regenerated with a high salt wash of 6 mL buffer B following the re-equilibration of the column with 10 mL buf- fer A. The proteins were detected at 280 nm and conductivity signal was recorded to monitor the salt gradient. CONCLUSION This straightforward application illustrates the principle of ion exchange chromatography. Two model proteins with different surface charges eluted under increasing salt concentrations from the cation exchange column. The proteins could be easily separated by cation exchange chromatography with the AZURA® Bio purification system. Fig. 1 Principle of cation exchange separation A) proteins with different negative charges bind to the cation exchange resin, B) by increasing the salt concentration proteins with a weak negative charge elute first, C) with higher salt concentrations proteins with a strong ne- gative charge elute last Fig. 2 Chromatogram of the separation of two model proteins with ion exchange chro- matography, blue line - UV 280 nm signal, red line - conductivity signal, 1) peak containing α-Chymotrypsinogen A (2.5 mg/mL), 2) peak containing Lysozyme (2.5 mg/mL)

VBS0065 Separation of two model proteins with Ion Exchange ... · Ion exchange chromatography is one of the most widely used FPLC techniques for protein separation and purification

  • Upload
    others

  • View
    4

  • Download
    0

Embed Size (px)

Citation preview

Page 1: VBS0065 Separation of two model proteins with Ion Exchange ... · Ion exchange chromatography is one of the most widely used FPLC techniques for protein separation and purification

Separation of two model proteins with Ion Exchange Chromatography (IEC) Michaela Schulze, Ulrike Maschke, Kate Monks; [email protected] Wissenschaftliche Geräte GmbH, Hegauer Weg 38, 14163 Berlin; www.knauer.net

SUMMARY

VBS0065© KNAUER Wissenschaftliche Geräte GmbHAdditional information

Ion exchange chromatography is one of the most widely used FPLC techniques for protein separation and

purification. This application describes an easy separation of two model proteins and explains how ion exchange

chromatography works.INTRODUCTIONIon exchange chromatography (IEC) separates molecules according to type and strength of their charge. The isoelectric point (pI) is the pH where a protein or molecule has no net elec-trical charge. Depending on the pH of the buffer a protein has different surface charges in solution. At a pH above their pI proteins have a negative charge and bind to positively charged resins such as anion exchangers. At a pH below their pI proteins have positive charge and bind to negatively charged cation exchangers. This interaction is used for the separation and purification of various proteins. By using a suitable pH and low salt conditions proteins bind to the resin in the initial step (Fig. 1A). The proteins are mostly separated with a linear salt gradient. The salt ions compete with the proteins for bindings sites. Proteins with weak ionic interactions are the first to elute from the column. In the case of cation exchange chromato-graphy, proteins that are less negatively charged start to elute first (Fig. 1B). With an increase of the salt concentration proteins with stronger ionic interaction elute later from the column (Fig. 1C).

RESULTSα-Chymotrypsinogen A (pI 8.97) and Lysozyme (pI 11.35) are proteins with rather high pI values, which make them ideal candidates for cation exchange chromatography (Fig. 2). Both proteins bind under low salt conditions to the resin. During the wash step that removes potentially unbound protein only a small amount of impurities eluted from the column. α-Chymot-rypsinogen A eluted first from the column due to its lower pI of 8.97 (Fig. 2, peak 1). With an increasing gradient and therefore increasing salt concentration lysozyme eluted as second peak (Fig. 2, peak 2). The eluted proteins were automatically collected by the fraction collector. The salt gradient was monitored by the conductivity monitor (Fig. 2, red signal).

MATERIALS AND METHODIn this application, an AZURA® Bio purification system consisting of AZURA P 6.1L LPG metal-free pump with 10 mL pump head; AZURA ASM 2.1L assistant module with feed pump and two injection valves; an AZURA DAD 2.1L diode array detector with 10 mm, 10 µL flow cell; AZURA CM 2.1S conductivity monitor and Foxy R1 fraction collector was used. Prior to the run the 1 mL cation exchange column (HiTrap Capto S) was equilibrated with 10 mL buffer A (20 mM phosphate buffer pH 7.1). The flowrate for the run was 2 mL/min. 500 µL protein mixture (α-Chymotrypsinogen A 2.5 mg/mL + Lysozyme 2.5 mg/mL) was injected. Subsequently the column was washed with 4 mL buffer A to remove all unbound protein. The two proteins were eluted with a linear gradient over a total volume of 70 mL. The salt concentration increased with the rising concentraion of buffer B (20 mM phosphate buffer pH 7.1 + 1M NaCl) in the gradient from 0 % up to 30 %. Eluting proteins were collected with the fraction collector. The column was regenerated with a high salt wash of 6 mL buffer B following the re-equilibration of the column with 10 mL buf-fer A. The proteins were detected at 280 nm and conductivity signal was recorded to monitor the salt gradient.

CONCLUSIONThis straightforward application illustrates the principle of ion exchange chromatography. Two model proteins with different surface charges eluted under increasing salt concentrations from the cation exchange column. The proteins could be easily separated by cation exchange chromatography with the AZURA® Bio purification system.

Fig. 1 Principle of cation exchange separation A) proteins with different negative charges bind to the cation exchange resin, B) by increasing the salt concentration proteins with a weak negative charge elute first, C) with higher salt concentrations proteins with a strong ne- gative charge elute last

Fig. 2 Chromatogram of the separation of two model proteins with ion exchange chro- matography, blue line - UV 280 nm signal, red line - conductivity signal, 1) peak containing α-Chymotrypsinogen A (2.5 mg/mL), 2) peak containing Lysozyme (2.5 mg/mL)

Page 2: VBS0065 Separation of two model proteins with Ion Exchange ... · Ion exchange chromatography is one of the most widely used FPLC techniques for protein separation and purification

ADDITIONAL MATERIALS AND METHODS

Separation of two model proteins with Ion Exchange Chromatography (IEC)

Eluent A 20 mM phosphate buffer pH 7.1

Eluent B 20 mM phosphate buffer pH 7.1 + 1 M NaCl

Gradient Volume [mL] % A % B

0 100 0

4 100 0

74 70 30

74.02 0 100

80 0 100

80 100 0

90 100 0

Flow rate 2 mL/min Run time 45 min

Column temperature 25°C Injection mode Full loop

Injection volume 0.5 mL Data rate 2 Hz

Detection wavelength 280 nm Time constant 500 ms

Tab. A1 Method parameters

Instrument Description Article No.

Pump AZURA P 6.1L LPG, metal-free APH64EB

Detector AZURA DAD 2.1L ADC01

Flow cell 10mm, 10µL, Ti, 300bar AMC38

Assistant

AZURA ASM 2.1L Left: Pump with pressure sensor, 50 mL pump head, titanium Middle: 6 port 2 position injection valve, 1/16" connectorsRight: 6 port 2 position injection valve, 1/16" connectors

AYBHECEC

Fraction collector Foxy R1 A59100

Conductivity monitor AZURA CM 2.1S ADG30

Column Capto S, GE Healthcare, HiTrap, 1mL -

Software PurityChrom 3D Option A2654

Tab. A2 System configuration & data

RELATED KNAUER APPLICATIONS

VBS0063 - Automated two - step purification of mouse antibody IgG1 with AZURA Bio LC Lab system

VBS0064 - Comparison of IgG purification by two different protein A media

VBS0067 - Automated two-step purification of 6xHis-tagged GFP with AZURA Bio LC

VBS0068 - Fast and robust purification of antibodies from human serum with a new monolithic protein A column

VBS0066 - Fast and sensitive size exclusion chromatography of IgG antibody

VBS0069 - Purification of Sulfhydryl Oxidase